Scope of application:
Used to measure the far-infrared emissivity of textiles.
Instrument features:
Place the standard blackbody plate and the sample on the hot plate in sequence, and adjust the surface temperature of the hot plate to reach the specified temperature; Using a far-infrared radiation measurement system with a spectral response range covering the 5 μ m-14 μ m wavelength band, the stable radiation intensity of a standard blackbody plate and a sample covered on a hot plate is measured. By calculating the ratio of the radiation intensity of the sample to that of the standard blackbody plate, the far-infrared emissivity of the sample is determined.
Applicable standards:
GB/T 30127(4.1)-2013
Technical Specifications:
Radiation source: 5-14um
Hot plate size: a circular surface with a diameter of not less than 60mm
Hot plate temperature: (34 ± 0.1) ℃
Test accuracy: ± 0.1%
Standard blackbody emissivity: above 0.95
Size: 390 × 335 × 320mm